Model | BNUV-R780 |
Optical system | Split beam,1200 l/mm grating |
Wavelength range | 190~1100 nm |
Spectral bandwidth | 1 nm |
Wavelength accuracy | ±0.3 nm |
Wavelength repeatability | ≤0.2 nm |
Wavelength resolution | 0.1 nm |
Wavelength swing speed | 10000 nm/min |
Wavelength scanning speed | 20 ~ 3200 nm/min |
Photometric range | 0 to 400%T, -4 to 4A, 0 to 9999C |
Photometric accuracy | ±0.3%T or ±0.004A at 1A |
Photometric repeatability | ±0.001 A(0~0.5 A),±0.002 A(0.5~1 A), ±0.15 %T(0~100 %T) |
Photometric Noise | ≤0.0002 A at 0 A,500 nm,≤0.0004 A at 1 A,500 nm,≤0.0008 A at 2 A,500 nm |
Photometric Stability | ≤0.0005A/h (500nm, 1 hours after preheating) |
Stray light | ≤0.05 %Tat 220nm & 360nm |
Baseline Flatness | ±0.001A |
Function | Photometry, Multi Wavelength, Kinetics, Time Scanning, Quantitation, Spectrum Scanning, DNA/Protein |
Light source | Tungsten lamp, Deuterium lamp |
Detector | Silicon photodiode |
Standard Sample holder | 10 mm 4-cell manual holder |
Display | 8'' inch TFT color touch screen |
Storage | 64GB |
Interface | USB-A × 1 (USB memory/USB printer), USB-B × 1 (computer),Wifi,SD Reader, Bluetooth,HDMI |
PC software | YES |
Language | 10 Languages, Can Be Switched (English, German, French, Spanish, Portuguese, Italian, Russian, Simplified Chinese, Japanese, Korean) |
Power supply | AC 100~240 V,50/60 Hz,140W |
Size | 456(W)×360(D)×185(H) mm |
Weight | 11 kg |
Split Double Beam Design
Bluetooth Wireless Connected Tablet PC with Windows operating system
8 inch high resolution capacitive Color Touch Screen
Optional software followed by GLP/GMP, Data traceable
Dismountable lamp room cover, can change lamp easily
The front baffle of the sample room can be dismantled freely
Special leak hole in the sample room
Antirust treatment cell holder.