Technical Parameters:
Chamber dimensions (LxWxH) mm | 650x380x260 |
Volume (L) | 64L |
Overall dimensions (LxWxH) mm | 850x725x750 |
Weight (kg) | 200 |
Power supply | 220V 50HZ |
Input power (kVA) | 3.5 |
Environment temperature | 10~40 |
Minimum installation room (LxW) mm | 1500x900 |
Minimum repair distance | 1 meter for each side |
Automatic Lifting Trolley:
Overall dimensions(LxWxH) mm | 1150x780x1165 | 1650x780x1165 |
Suitable sterilizer | PS-100X / PS-100XP | PS-200X / PS-200XP |
Low temperature sterilization
The sterilization temperature is 50±5℃ which is much lower than steam sterilization. It is dry sterilization that has no harm on instrument, so it extends the device service life.
Safe and environment friendly
The matter after sterilization is water and oxygen molecule that have no toxic harm to staff or patient, and there is no need for drainage or air exhausting. High efficient H2O2 fifilter is installed in the exhausting outlet to guarantee no H2O2 pollution to the environment.
Time-saving and cost-efficient
Rectangular chamber has greater capacity, high efficiency and short sterilizing cycle (≤50mins). The sterilized articles can be used directly without extra processing. It can increase the instrument turnover and decrease the storage of precious instrument to cut cost indirectly.
Convenient
Convenient
Only power supply is needed for installation. The system adopts man-machine control interface with intelligent program design.
Equipped with foot switch (alternative) and integrated biological incubator (alternative)
Multiple models
The door structure of the sterilizer is customizable (single or double door). To meet pharmaceutical factory needs, chamber material can be stainless steel.
Reliable sterilization
The chamber adopts alloy material which has good thermal conductivity to guarantee a stable temperature in the chamber. The purifification system makes the H2O2 diffuses uniformly. The low temperature storage system provides a stable concentration of H2O2
Safe sealing door
The sterilizer adopts mandril-driving auto overhead door with reliable sealing function and safe lock device to guarantee personal safety.
Advanced confifiguration
The vacuum pump, controller, pressure transmitter and other key components are world famous brand. The sterilizer equipped with multiple programs and bar code identifification system (alternative)
Reassuring
We provide comprehensive operation training materials to key components maintenance intelligent reminding system for your convenience.
The sterilizer is applicable to sterilize the medical instruments that are thermolabile or non-moisturetight. It is not applicable to the sterilization of liquid, powder or cotton fabrics. All the medical instruments that meet the following conditions can be processed by this sterilizer.
BN-T-100:
Metal tubular instrument: inner diameter ≥ 1mm; length ≤500mm
Nonmetal tubular instrument: inner diameter≥ 1 mm; length≤2000mm
BN-T-200:
Metal tubular instrument: inner diameter≥0.8mm; length≤300mm
Nonmetal tubular instrument: inner diameter≥1 mm; length≤1000mm