Model | BBSC-X1500IIB2 | |
External Size(W*D*H) Internal Size(W*D*H) |
1500*760*2250mm 1350*600*660mm |
|
Tested Opening Max Opening |
Safety Height 200mm(8") | |
500mm(20") | ||
Inflow Velocity Downflow Velocity |
0.5±0.025 m/s 0.3±0.025 m/s |
|
Pre-Filter HEPA Filter Front Window Noise |
Washable Two, 99.999% efficiency at 0.3µm. Filter life indicator. Motorized. Two-layer laminated toughened glass≥5mm. Anti UV. EN12469≤58dB / NSF49≤ 61dB |
|
UV Lamp | 40W*1 | |
UV timer, UV life indicator, emission of 253. 7 nanometers for most efficient decontamination. | ||
Illuminating Lamp | LED lamp | |
16W*2 | ||
Illumination Consumption |
≥1000Lux | |
900W | ||
Waterproof Sockets | Two, total load of two sockets: 500W | |
Display | LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow Display velocity, filter life, humidity and temperature, system working time etc. | |
Control System Airflow System Visual and audio alarm |
Microprocessor 0% air recirculation, 100% air exhaust Abnormal airflow velocity; Filter replacement; Front window at unsafe height. |
|
Material | Work Zone: 304 stainless steel Main Body: Cold-rolled steel with anti-bacteria powder coating. |
|
Work Surface Height Caster |
750mm Foot master caster |
|
Power Supply | AC220V±10%, 50/60Hz; 110V±10%, 60Hz | |
Standard Accessory Optional Accessory |
Illuminating lamp, UV lamp*2, Base stand, Remote control, Foot switch, Exhaust blower, Exhaust Duct, Drain Valve, Waterproof sockets*2, pipe strap*2 | |
Gross Weight | 302kg | |
Package | Main Body Exhaust Blower( W*D*H ) |
1650*990*1810mm 970*810*630mm |
Time reserve function.
Motorized front window.
HEPA fifilter life and UV life indicator.
Automatic air speed adjustable with fifilter block.
With memory function in case of power-failure.
Work area surrounded by negative pressure, it can ensure maximum safety in work area.
Most accessories are standard, which saves your money. There is no need to pay more.
Audio and visual alarm (fifilter replacement, window over height, abnormal air flflow velocity, etc.).
Remote control. All functions can be realized with it, making the operation much easier and more convenient.
Interlock function: UV lamp and front window; UV lamp and blower, flfluorescent lamp; blower and front window
Foot switch. Adjust front window height by foot during experiment, to avoid airflflow turbulence caused by arm movement.