Model | BBSC-X1300IIA2 |
External Size(W*D*H) Internal Size(W*D*H) |
1300*750*2250mm 1150*600*660mm |
Tested Opening Max Opening |
Safety Height 200mm(8") |
420mm(17") | |
Inflow Velocity Downflow Velocity |
0.53±0.025 m/s 0.33±0.025 m/s |
HEPA Filter Front Window Noise |
Two, 99.999% efficiency at 0.3µm. Filter life indicator. Motorized. Two-layer laminated toughened glass≥5mm. Anti UV. EN12469≤58dB / NSF49≤ 61dB |
UV Lamp | 30W*1 |
UV timer, UV life indicator, emission of 253. 7 nanometers for most efficient decontamination. | |
Illuminating Lamp | LED lamp |
12W*2 | |
Illumination Consumption |
≥1000Lux |
800W | |
Waterproof Sockets | Two, total load of two sockets: 500W |
Display | LCD display: exhaust filter and downflow filter pressure, filter and UV lamp working time, inflow and downflow Display velocity, filter life, humidity and temperature, system working time etc. |
Control System Airflow System Visual and audio alarm |
Microprocessor 70% air recirculation, 30% air exhaust Abnormal airflow velocity; Filter replacement; Front window at unsafe height. |
Material | Work Zone: 304 stainless steel Main Body: Cold-rolled steel with anti-bacteria powder coating. |
Work Surface Height Caster |
750mm Foot master caster |
Power Supply | AC220V±0%, 50/60Hz; 11 OV±10%, 60Hz |
Standard Accessory Optional Accessory |
LED lamp, UV lamp*2, Base stand, Remote control, Foot switch, Drain valve, Waterproof sockets*2 Water and gas tap, Electric height adjustable base stand |
Gross Weight Package Size(W*D*H) |
280kg 1460*1050*1800mm |
Motorized front window.
Time reserve function.
HEPA fifilter life and UV life indicator.
Automatic air speed adjustable with fifilter block.
With memory function in case of power-failure.
Audio and visual alarm (Abnormal airflflow velocity; Filter replacement; Front window at unsafe height).
Work area surrounded by negative pressure, it can ensure maximum safety in work area.
Audio and visual alarm (Filter replacement, window over height, abnormal air flflow velocity etc.).
Remote control: all functions can be realized with it, making the operation much easier and more convenient.
Interlock function: UV lamp and front window; UV lamp and blower,
LED lamp; blower and front window.
Foot switch: adjust front window height by foot during experiment, to avoid airflflow turbulence caused by arm movement.
No PDF document available for this product at present
Leave a message
Please fill in information, we will get back to you within 24 hours.